黄色一级视_久久丫精品国产免费_欧美va免费高清在线观看_yw855.ccm免费观看网址_MM1313亚洲精品无码久久_97超碰色

Fluorescence Fiber Optic Temperature On-line Monitoring Scheme For Medical Industry

Fluorescence Fiber Optic Temperature On-line Monitoring Scheme For Medical Industry

Technical Background

The temperature treatment methods in medical industry are microwave therapy, microwave hyperthermia, radiofrequency hyperthermia, and magnetic resonance imaging. In microwave hyperthermia, the main factor affecting the accuracy of temperature measurement results is the interference of microwave or radio frequency. The metal made temperature sensor probe will produce inductive current in microwave field, which makes the measurement results cannot reflect the real temperature of tumor tissues. Whether hyperthermia is effective, the key question is to heat the tumor above the effective treatment temperature. If the tumor temperature is below the treatment temperature, not only has no treatment effect, but also may promote the spread of the tumor, if the temperature is too high, may damage the normal tissue, in about 42℃,1℃ difference can cause multiple changes of cell survival rate. Clinical temperature measurement accuracy within 37℃ ~ 46℃ should be within ±0.2℃, so accurate temperature measurement is the key technology in microwave hyperthermia. 

Comparison of Temperature Measurement Methods


Index


Fluorescence Fiber Optic Temperature Sensor


Conventional Temperature Sensor


Thermistor- High Resistance Wire Temperature Sensor


Operating environment


High voltage under strong electromagnetic field



High voltage under strong electromagnetic field



High voltage under strong electromagnetic field



Materials


No metal parts, using fiber optic sensor probe


Metal made temperature probe and wire


Employ semiconductor thermistors as probes, high resistance wire as signal transmission line, and equipped with a simple measurement circuit


Advantages and disadvantages



1.Anti-electromagnetic interference, high voltage resistance, corrosion resistance

2.Reliable and stable, high precision, anti-explosion and flame-proof

3. Tiny size and light weight

4. Non-toxic


1. Under high frequency electromagnetic field, the temperature probe and wire will produce induced current, which increases theirs’ temperature and interferes with temperature.

2. Big error makes it impossible to measure stable temperature.

3.Also bring great harm to microwave equipment


1. Poor stability, interchangeability and linearity

2.Poor mechanical strength of high resistance wire, difficult to achieve high accuracy measurement


Under the special environment of microwave, radio frequency and high frequency, the fluorescence fiber temperature measurement system can measure the online temperature in real time, which exactly reflects the real temperature of the measured position, and is the optimal temperature measurement method. 

HQ-16/18 Fluorescence Fiber Optic Temperature Monitoring System

Product Description

Name

Performance

Picture

One channel temperature measuring system-

rectum treatment


Reading temperature range: 0-80℃Resolution:0.1℃

Temperature accuracy: ±0.3℃

Sensing cable length:< 10m   

Sensing cable diameter:2.3mm

Input power: DC24V

Communication port:485/RS-232Modbus

(protocol / RdLight Logger software)

Analog output:4-20mA or 0-5V

Power consumption:1W

One channel temperature measuring system-Tumor hyperthermia and microwave therapy


Reading temperature range: 30-50℃Resolution:0.1℃

Temperature accuracy: ±0.3℃

Sensing cable length:< 10m   

Sensing cable diameter:0.45mm

Input power: DC24V

Communication port:485/RS-232Modbus

(protocol / RdLight Logger software)

Analog output:4-20mA or 0-5V

Power consumption:1W

Six channels temperature measuring system

Reading temperature range: 30-50℃Resolution:0.1℃

Temperature accuracy: ±0.3℃

Sensing cable length:< 10m   

Sensing cable diameter:0.45mm

Input power: 220V

Communication port:485/RS-232Modbus

(protocol / RdLight Logger software)

Analog output:4-20mA or 0-5V

Power consumption:1W

Software Introduction

The supporting software of fluorescence fiber optic temperature measurement products have the functions of reading, storage, drawing and data export, which are convenient for real-time observation and data post-processing.

Temperature Profile

Data Storage Management


主站蜘蛛池模板: 九九久久精品一区二区三区 | 国产精品久久久久久久久久直播 | 干女人网站 | 亚洲3atv精品一区二区三区 | 国产精品视频在线看 | 亚洲国产av无码精品无广告 | 二级性生活片 | 中国白嫩的sex少妇黄网免费 | 91久久久精品国产 | 撕开奶罩揉吮奶头免费视频 | 美国一级黄色片 | 手机av在线网址 | 娇小12一13sexvideo | 天天射天天射 | 小鸟酱福利视频 | 干女人网站 | 一本大道大臿蕉无码视频 | 自拍偷区亚洲综合美利坚 | 成人无码在线视频区 | 日本av在线播放 | 美女写真理伦片在线看 | 野花香日本大全免费观看 | 日本熟妇厨房XXXXX乱 | 亚洲AV永久无码一区二区三区 | 成人免费视频xbxb入口 | 少妇激情偷公乱柔佳 | 日韩福利一区二区 | 国产一卡二卡新区在线 | 99久久夜色精品国产亚洲av卜 | 蜜桃成人无码区免费视频网站 | 91碰在线视频 | 国产人妻久久精品二区三区 | 97国产精华最好的产品亚洲 | 用力插h | 99成人在线 | 久久99精品视频一区97 | 国产又色又爽又高潮免费视频麻豆 | yellow中文字幕国产资源 | 欧美三级小视频 | 3344永久在线观看视频免费 | 成人国产一区二区 |